Configure Item
Duration Time
4.5.2 ID Input
Press soft key "ID" to enter configuration screen after the average weight is displayed.
Press the ID key, and you can see there is a window for you to type the ID number.
Press the ID key again to save the ID number.
4.5.3 Average Time Setup
There are two methods to set the averaging time.
The default Average Time is 10 seconds.
Note: When the time is set to 0, the first stable weight over 5d will be displayed.
Averaging time can be set to a value between 0 and 20 seconds.
Find Quality Products Online at:
Option(Bold is default)
1 ~ 10 s
GlobalTestSupply
www.
Defender 5000 Indicators
Description
Automatic: The Softkey "∑" displays. The weight will
be accumulated automatically.
Note:
1. The load to be accumulated has to be >= 10d.
Another accumulation can't be done until the
pan is cleared (< 5d).
2. Gross weight and net weight can't be
accumulated together when the LFT is ON (no
such limit when the LFT is OFF or the
approved model is OIML). When first
accumulated value is gross weight (net
weight), scale will enter into gross weight (net
weight) accumulate mode. Otherwise, "Gross
and net weight can't be accumulated" will be
displayed.
3. If the LFT is ON, gross weight 0 has to be
reached before a new sample can be
accumulated. If the LFT is OFF, new sample
can be accumulated after reaching gross
weight 0 or net weight 0.
It is the time for the display to remain the dynamic
weighing result after the load is removed.