EN-18
3.3 Calibration Menu
The TD52 indicator offers three calibration methods:
Zero Calibration, Span Calibration and Linearity Calibration.
NOTES:
1. Make sure that appropriate calibration masses are available before calibration.
2. Make sure that the scale base is level and stable during the entire calibration process.
3. Calibration is unavailable with LFT set to ON.
4. Allow the Indicator to warm up for approximately 5 minutes after stabilizing to room temperature.
5. To abort calibration, press the Softkey corresponding to the icon 'X' anytime during the calibration
process.
6. When any selection within the GMP menu is enabled, calibration results are automatically printed.
7. When the calibration is completed, the output of the calibration result is in the same unit as the
weighing status.
3.3.1 Zero Calibration
Zero calibration uses one calibration point. The zero calibration point is established with no weight
on the scale. Use this calibration method to adjust for a different dead load without affecting the span
or linearity calibration.
Calibration procedures:
Long press the button
Main Menu. Press the Softkey corresponding to
the icon
Zero Calibration is on the top of the list of
Calibration by default. Just press the Softkey
corresponding to the icon
Calibration.
Clear the pan and then press the Softkey
corresponding to the icon
The message 'Completed' will be displayed on
the screen.
Exit Zero Calibration by pressing the Softkey
corresponding to the icon
To return to the Main Menu, press the Softkey
corresponding to the icon
3.3.2 Span Calibration
Span Calibration uses one point. The span calibration point is established with a calibration mass
placed on the scale.
Note: Span Calibration should be performed after Zero Calibration.
Find Quality Products Online at:
to enter the
to enter the Calibration sub-menu.
to initiate Zero
.
.
.
GlobalTestSupply
www.
Defender 5000 Indicators
.com
sales@GlobalTestSupply.com