EN-12
or the individual wires can be wrapped through the ferrite. This should be done as close to the
enclosure as possible. See Figure 2-2.
Main Board Wiring Connections
Once the TD52P and TD52XW enclosure is open, connections can be made to the terminal blocks on
the main board, as shown in Figure 2-3.
Jumper Connections
The TD52P and TD52XW indicators are designed to support both 2mV/V and 3mV/V load cells from
the same circuitry. A load cell output rating selection jumper is not required.
Figure 2-4 shows the terminal definitions for the analog load cell terminal blocks. Note that when using
four-wire load cells, jumpers must be placed between the +Excitation and +Sense terminals and
between the Excitation and Sense terminals.