EN-42
4.2.2 Application Setup
The Application can be customized for various user preferences.
Press the Softkey corresponding to the icon
The Configuration screen is now displayed.
Select the list item and press the Softkey
corresponding to the icon
setting as desired.
To return to the Application home screen, press
the button corresponding to the icon
The Counting Configurations are defined below (defaults in Bold).
Configure Item
Auto Tare
Accumulate
Auto Opt.
Opt. Beep
APW Auto Save
Internal Resolution
Reference Size
4.2.3 Accumulation
See section 4.1.3 for details about the Accumulation feature.
Find Quality Products Online at:
.
to enter Configuration
to change the
Option(Bold is
default)
Off/On
Off/
Manual/
Automatic
Off/On
Off/On
Off/On
Off/On
10
GlobalTestSupply
www.
.
Description
Off: Auto tare is turned off.
On: The first stable weight (>=5d) will be tared as a container
automatically.
Off: The icon "∑" does not display.
Manual: The Softkey "∑" displays. User can press the key to
accumulate the stable weight.
Automatic: The icon "∑" displays. The weight will be accumulated
automatically.
Note:
The load to be accumulated has to be >= 5d and the next accumulation
can only start once the pan has been cleared.
no such limit when LFT is OFF or the approve
When LFT is ON (
mod is OIML
),
a. gross weight and net weight cannot be accumulated;
b. gross 0 has to be reached before a new sample accumulation.
When LFT is OFF,
a. gross weight and net weight can be accumulated;
b. a new sample can be accumulated after reaching gross 0 or net 0.
Off: Auto Opt. is off.
On: The APW will be optimized automatically during count weighing.
Off: Opt. Beep is off.
On: When the APW has to be optimized, the beeper will beep once.
Off: APW Auto Save is off.
On: If the APW is derived from sampling, and a counting library is
selected, the new APW will be saved to library after optimized.
Note: It will be hidden when "Auto Opt." is off.
Off: Internal Resolution is off.
On: During sampling or weighing, the internal resolution will be used.
The One Button Sampling PCS is from 0 to 999, the default is 10.
0: The One Button Sampling key will be hidden.
Defender 5000 Indicators
.com