EN-22
When Dual interval is selected, the terminal functions with two intervals, each with its own capacity
and graduation. In addition to the Interval 1 capacity and graduation parameters, the following two
parameters are available:
•
•
3.4.3 Capacity
Set the scale capacity from 1 to 999999.
>|1|< Capacity
Specify the weight capacity for interval 1. If Single interval is enabled, this will be the scale capacity.
If Dual interval is enabled, this will be the first range.
>|2|< Capacity
Specify the weight capacity for interval 2. If Dual interval is enabled, this will be the scale capacity
and it must be bigger than >|1|< Capacity. If Single interval is enabled, this parameter will not be
shown.
3.4.4 Graduation
Set the scale readability from 0.0001 to 100.
>|1|<Graduation
Specify the graduation for weighing interval 1. If Single interval is enabled, this will be the graduation
for the entire weighing range of the scale. If Dual interval is enabled, this will be the graduation used
in the lower interval.
>|2|<Graduation
Specify the graduation for interval 2. If Dual interval is enabled, this will be the graduation for the
second weighing range of the scale. If Single interval is enabled, this parameter will not be shown.
NOTE: Graduation settings are limited to values from Capacity divided by 600 to Capacity divided
by 75000. Therefore, not all settings are available for each capacity.
3.4.5 Language
Set the language displayed for menus and displayed messages.
English
Deutsch
Français
Italiano
Polski
Spanish
한국
中文
日本语
3.4.6 Power On Zero
Zero the balance at Power On.
When the power on zero function is turned off, the user performs the tare operation. The terminal will
reuse the tare value after the terminal is restarted.
Off = disabled.
On = enabled.