Defender 5000 Indicators
Calibration procedures:
Long press the button
Menu.
Press the Softkey corresponding to the icon
enter the Calibration sub-menu.
Scroll to Span Calibration using the Softkey
corresponding to the icon
Press the Softkey corresponding to the icon
initiate Span Calibration.
Place a calibration mass of the specified weight on
the pan and press the Softkey corresponding to the
icon
input the value desired, and then place the
corresponding weight on the pan for calibration.
A suggestive message shows on the screen.
The message 'Completed' will be displayed on the
screen.
Exit Span Calibration by pressing the Softkey
corresponding to the icon
To return to the Main Menu, press the Softkey
corresponding to the icon
Note: Span Calibration should be performed after
Zero Calibration.
3.3.3 Linearity Calibration
Linearity calibration uses 3 calibration points. The full calibration point is established with a weight
on the scale. The mid calibration point is established with a weight equal to half of the full calibration
weight on the scale. The zero calibration point is established with no weight on the scale. The full
calibration and mid calibration points can be altered by the user during the calibration procedure.
Calibration procedures:
Long press the button
Menu.