****CALIBRATION DATA****
3/18/19
13:54:30
SERIAL #
030021901
**CURRENT CALIBRATION**
Chamber Pressure Low
PT1 Sensor: 10V
3/12/19
14:45:38
Cal. Name:
BR
Cal type: Edit
LOW RAW = 8390937
HIGH RAW = 12790467
LOW VALUE = 0.10 Torr
HIGH VALUE = 23.000 Torr
**CURRENT CALIBRATION**
Chamber Pressure High
PT2 Sensor: 4 to 20
3/12/19
14:45:40
Cal. Name:
BR
Cal type: Edit
LOW RAW = 2834218
HIGH RAW =11334934
LOW VALUE = 0.2 Torr
HIGH VALUE = 745.9 Torr
**CURRENT CALIBRATION**
Vaporizer Temperature
RTD1 SENSOR: RTD
3/12/19
14:45:45
Cal. Name:
BR
Cal type: Edit
LOW RAW = 9975924
HIGH RAW = 11913543
LOW VALUE = 23.0-C
HIGH VALUE = 74.9 -C
**CURRENT CALIBRATION**
Vaporizer Temperature
RTD2 SENSOR: RTD
3/12/19
14:45:47
Cal. Name:
BR
Cal type: Edit
LOW RAW = 9992758
HIGH RAW = 11934117
LOW VALUE = 23.0-C
HIGH VALUE = 74.9 -C
**CURRENT CALIBRATION**
Chamber Temperature
RTD3 SENSOR: RTD
3/12/19
14:45:49
Cal. Name:
BR
Cal type: Edit
LOW RAW = 9983903
HIGH RAW = 11919835
LOW VALUE = 23.0-C
HIGH VALUE = 74.9 -C
**CURRENT CALIBRATION**
Chamber Temperature
RTD4 SENSOR: RTD
3/12/19
14:45:53
Cal. Name:
BR
Cal type: Edit
LOW RAW = 10019474
HIGH RAW = 11959734
LOW VALUE = 23.0-C
HIGH VALUE = 74.9 -C
Système de stérilisation à basse
température
9 — Rapports imprimés