Internal Battery
Housing construction
Environmental
conditions
Electromagnetic
compatibility
Air filter
Air tubing
SlimLine air tubing
IEC 60601-1
classifications
Air travel requirements
This device is not suitable for use in the presence of a flammable anesthetic mixture.
Notes:
• The manufacturer reserves the right to change these specifications without notice.
• Pressure may be displayed in cm H
Lithium-Ion battery, 14.4 V, 1.6 Ah, 23 Wh
Operating hours: 2 h with a new battery under normal
conditions (see below).
Patient type: home chronic; pressure: IPAP/EPAP 15/5
cm H
O; mask type: Ultra Mirage; air tubing: 2 m; leak:
2
0; respiratory rate: 20 bpm; battery capacity: 100%
Patient type: hospital acute; pressure: IPAP/EPAP 20/5
cm H
O; mask type: Ultra Mirage; air tubing: 2 m; leak:
2
0; respiratory rate: 45 bpm; battery capacity: 100%
Flame retardant engineering thermoplastic
• Operating temperature: 0°C to 35°C
• Operating humidity: 10%–95% non-condensing
• Storage and transport temperature: -20°C to 60°C
(+50°C*)
• Storage and transport humidity: 10%–95% non-
condensing
• Air pressure: 680hPa to 1,100hPa; Altitude: 3,000 m
*NONIN XPOD
Product complies with all applicable electromagnetic
compatibility requirements (EMC) according to
IEC60601-1-2, for residential, commercial, and light
industry environments. For further details see "Guidance
and manufacturer's declaration – electromagnetic
emissions and immunity" on page 32.
Electro static fibre mesh with TPE frame structure. Bacterial
filtration efficiency of 99.540% on area weight 100g/m².
Flexible plastic, 2 m or 3 m length (22 mm diameter)
Flexible plastic, 1.83 m length (15 mm diameter)
• Class II (Clause 3.14—double insulation). This
adherence means the need for an protective earthing
(ie, an earthed plug) is not necessary.
• Type BF
• Continuous operation
Medical-Portable Electronic Devices (M-PED) that meet
the Federal Aviation Administration (FAA) requirements
of RTCA/DO-160 can be used during all phases of air
travel without further testing or approval by the airline
operator. ResMed confirms that the Stellar meets RTCA/
DO-160 requirements.
O or hPa.
2
Technical specifications
31