Technical specifications
Operating pressure
range
Maximum single fault
pressure
Maximum breathing
resistance under single
fault
Maximum flow
Flow accuracy
Therapy pressure
tolerance
Sound pressure level
Sound power level
Alarm Volume Range
Dimensions (L x W x H)
Weight
LCD screen
Air outlet
Pressure measurement
Flow measurement
Power supply
External DC Power
Supply (isolated)
30
• IPAP: 2 cm H
O to 40 cm H
2
• PS: 0 cm H
O to 38 cm H
2
• EPAP: 2 cm H
O to 25 cm H
2
mode)
• CPAP: 4 cm H
O to 20 cm H
2
• Min PS: 0 cm H
O to 20 cm H
2
• Max PS: 0 cm H
O to 30 cm H
2
30 cm H
O when AutoEPAP is on (in iVAPS mode)
2
• Min EPAP/Max EPAP: 2 cm H
iVAPS mode when AutoEPAP is set to on)
Note: iVAPS mode is only available in Stellar 150.
60 cm H
O (in all modes)
2
2 cm H
O at 30 L/min;
2
7 .2 cm H
O at 60 L/min
2
> 200 L/min at 20 cm H
± 5 L/min or 20% measured value, whichever is greater
Test condition: T mode, IPAP: 40 cm H
H
O, Rise Time: MIN, Fall Time: MIN, Ti: 4.0 sec,
2
Respiratory Rate: 10 bpm, with ResMed calibration cap.
IPAP: ± 0.5 cm H
O ± 10% of set pressure (end of
2
inspiration)
EPAP/PEEP: ± 0.5 cm H
CPAP: ± 0.5 cm H
O ± 10% of set pressure
2
Test condition: T mode, IPAP: 40 cm H
H
O, Rise Time: MIN, Fall Time: MIN, Ti: 4.0 sec,
2
Respiratory Rate: 10 bpm, with ResMed calibration cap.
31 dBA with uncertainty of 3 dBA as measured according
to ISO 17510 – 1:2007 .
39 dBA with uncertainty of 3 dBA as measured according
to ISO 17510 – 1:2007 .
High, medium and low priority alarm (measured according
to IEC 60601-1-8: 2012):
minimum 41 dB
maximum 67 dB
230 mm x 170 mm x 120 mm
2.1 kg
Dimension (L x W x H): 76.9 mm x 63.9 mm x 3.15 mm
Resolution: 320 x 240 pixels
22 mm taper, compatible with ISO 5356-1:2004
Anaesthetic & Respiratory Equipment - Conical
Connectors
Internally mounted pressure transducer
Internally mounted flow transducer
AC 100–240V, 50–60Hz, 2.2 A, max. 65 W
24 V, 3 A
O (in S, ST, T, PAC mode)
2
O (in S, ST, T, PAC mode)
2
O (in S, ST, T, iVAPS, PAC
2
O (in CPAP mode only)
2
O (in iVAPS mode)
2
O; 8 cm H
2
2
O to 25 cm H
2
O
2
O, EPAP: 2 cm
2
O ± 4% of set pressure
2
O, EPAP: 2 cm
2
O to
O (in
2