Use of accessories, transducers, and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
To ensure compliance with the EMC requirements as specified in IEC 60601-1-2, it is
essential to use only the following accessories.
Item
IP30 Insert Earphone
DD45 Headphones
DD450 Headphones
DD65 v2 Headphones
B81 Bone Vibrator
Conformance to the EMC requirements as specified in IEC 60601-1-2 is ensured if the cable
types and cable lengths are as specified below:
Description
IP30 Insert Earphone
DD45 Headphones
DD450 Headphones
DD65 v2 Headphones
B81 Bone Vibrator
Power Supply Cable
USB Cable
NOTICE: The use of the accessories, transducers, and cables with medical
equipment/system other than this equipment may result in increased emissions or
decreased immunity of the medical equipment/system.
D-0131088 Rev C, 2022-07
GSI AudioStar Pro Instructions for Use
Manufacturer
RadioEar
RadioEar
RadioEar
RadioEar
RadioEar
Length
2.5 m
2.5 m
2.5 m
2.5 m
2 m
2.5 m
2 m
Item #
8101884
8503851
8520321
8510115
8501119
Screened/Unscreened
Unscreened
Unscreened
Unscreened
Unscreened
Unscreened
Unscreened
Screened
Page 84 of 84