G
M
UIDANCE AND
ANUFACTURER
E
LECTROMAGNETIC
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations
on power supply
lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
D-0131088 Rev C, 2022-07
GSI AudioStar Pro Instructions for Use
'
D
S
ECLARATION
I
MMUNITY
IEC 60601 Test level
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV differential
mode
±2 kV common mode
< 5% UT
(>95% dip in UT) for
0.5 cycle
40% UT
(60% dip in UT) for 5
cycles
70% UT
(30% dip in UT) for 25
cycles
<5% UT
(>95% dip in UT) for 5
sec
3 A/m
Compliance
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Not applicable
±1 kV for
input/output lines
Not applicable
Not applicable
3 A/m
Electromagnetic
Environment-Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be greater than 30%.
Mains power quality
should be that of a typical
commercial or residential
environment.
Mains power quality
should be that of a typical
commercial or residential
environment.
Mains power quality
should be that of a typical
commercial or residential
environment. If the user of
the system requires
continued operation
during power mains
interruptions, it is
recommended that the
system be powered from
an uninterruptable power
supply or its battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or residential
environment.