Test Procedure
Select QuickSIN from the More Tests Menu.
•
Ensure the proper transducer and intensity levels are selected.
•
Select the appropriate word list. Lists 1-12 are standard lists used to determine SNR
•
loss.
Using the Word Nav and front panel navigation buttons or an external mouse, select
•
the first sentence.
•
Press the present bar or click the selected sentence.
•
Score the five key words highlighted in each sentence by pressing the INCORRECT
button for each word repeated incorrectly by the patient.
Press Store
•
The SNR Loss score will appear in the SCORE/WORD window.
•
NOTE: It is recommended that at least 2 lists be presented in each condition. The
average score will appear in the Group 1 SNR averages window.
To interpret the SNR loss score, see the table above in the BKB-SIN section.
Refer to the QuickSIN user manual for advanced testing options.
D-0131088 Rev C, 2022-07
GSI AudioStar Pro Instructions for Use
Page 36 of 84