Test Procedure
Select QuickSIN from the More Tests Menu.
Ensure the proper transducer and intensity levels are selected.
Select the appropriate word list. Lists 1-12 are standard lists used to determine SNR
loss.
Using the Word Nav and front panel navigation buttons or an external mouse, select
the first sentence.
Press the present bar or click the selected sentence.
Score the five key words highlighted in each sentence by pressing the INCORRECT
button for each word repeated incorrectly by the patient.
Press Store
The SNR Loss score will appear in the SCORE/WORD window.
NOTE: It is recommended that at least 2 lists be presented in each condition. The
average score will appear in the Group 1 SNR averages window.
To interpret the SNR loss score, see the table above in the BKB-SIN section.
Refer to the QuickSIN user manual for advanced testing options.