Immunity test
IEC / EN 60601
test level
Conducted RF
3 Vrms
IEC / EN 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC / EN 61000-4-3
80 MHz to 2,5 GHz
D-0131088 Rev C, 2022-07
GSI AudioStar Pro Instructions for Use
Compliance level
3 Vrms
3 V/m
Electromagnetic
environment – guidance
Portable and mobile RF
communications
equipment should be used
no closer to any parts of
the system, including
cables, than the
recommended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended separation
distance
1
2 ,
d
=
P
d
=
1
2 ,
P
80 MHz to
800 MHz
d
2
3 ,
P
=
800 MHz to
2,5 GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the
transmitter manufacturer
and d is the recommended
separation distance in
meters (m).
Field strengths from fixed
RF transmitters, as deter-
mined by an
electromagnetic site
survey, (a) should be less
than the compliance level
in each frequency range (b)
Page 82 of 84