NOTE: If you select a different transducer or stimulus or ear and leave the tone test type,
when you return to tone, the AudioStar Pro will default to the last settings that were
selected.
Press the Tone Test Type Button.
•
•
Verify that the transducers and signals are correct.
•
Verify the frequency range is correct.
•
Perform air conduction threshold testing.
Press "store" after each threshold is obtained.
•
When the pure tone evaluation is complete, move to the next test type in your typical
testing sequence.
H
F
T
IGH
REQUENCY
Use the navigation buttons or the mouse to access the Range menu. Available frequency
ranges are Standard (125-8,000 Hz), High (8,000-20,000 Hz) or Full (125 – 20,000 Hz). Auto
Hz can be defined in the Config App and automatically moves the next designated test
frequency when "store" is pressed.
•
Press the Tone Test Type Button.
Select the desired frequency range (High or Full).
•
Verify that the transducers and signals are correct.
•
Perform High Frequency Testing.
•
Press "store" after each threshold is obtained.
•
When the high frequency evaluation is complete, move to the next test type in your typical
testing sequence
TEN T
T
EST
YPE
Presentation Level
If the audiometric threshold is 60 dB or less, set the TEN level to 70 dB.
•
If the audiometric threshold is 65 dB or greater, set the TEN level 10 dB higher than
•
the threshold.
If the TEN noise is reported to be too loud, set the TEN level at 90 dB (maximum) or
•
the same level as audiometric threshold.
D-0131088 Rev C, 2022-07
GSI AudioStar Pro Instructions for Use
ESTING