Portable and mobile RF communications equipment can affect the AudioStar Pro. Install
and operate the system according to the EMC information presented in this chapter.
The AudioStar Pro has been tested for EMC emissions and immunity as a stand-alone
device. Do not use the system adjacent to or stacked with other electronic equipment. If
adjacent or stacked use is necessary, the user should verify normal operation in the
configuration.
The use of accessories, transducers, and cables other than those specified, except for
servicing parts sold by GSI as replacement parts for internal components, may result in
increased EMISSIONS or decreased IMMUNITY of the device.
Anyone connecting additional equipment is responsible for making sure the system
complies with the IEC 60601-1-2 standard.
C
AUTIONS REGARDING
This instrument is suitable in hospital environments except for near active HF
surgical equipment and RF shielded rooms of systems for magnetic resonance
imaging, where the intensity of electromagnetic disturbance is high.
Use of this instrument adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this instrument and
the other equipment should be observed to verify that they are operating normally.
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the
AudioStar Pro, including cables specified by the manufacturer. Otherwise, degradation of
the performance of this equipment could result.
The AudioStar Pro is intended for use in the electromagnetic environment specified below.
The customer or the user of the Instrument should assure that it is used in such an
environment.
D-0131088 Rev C, 2022-07
GSI AudioStar Pro Instructions for Use
C
EMC
(EMC)
OMPATIBILITY