PERATING THE
P
C
RELIMINARY
HECKS
Before starting any procedures using the GSI AudioStar Pro Clinical Audiometer, ensure
that the power cord is plugged into a properly grounded receptacle.
Check also that all cords from the transducers, patient response hand switch
(if used), and printer fit securely in their connectors on the rear and side
panels. Inspect all cords for fraying and damage. If there is any damage to any
cord, do not use the AudioStar Pro. If speech testing with recorded voice from
an external source is to be performed, check that the CD or digital music player device is
connected and operating properly.
It is recommended that all parts that come into direct contact with the patient (e.g.,
earphone cushions) are subjected to standard disinfecting procedures between patients.
This includes physically cleaning and using a recognized disinfectant. Individual
manufacturer's instructions should be followed for use of any disinfecting agent to provide
an appropriate level of sterilization.
Handle earphones, bone vibrator, and insert earphones with care. Do not
drop them nor allow them to be banged together. Severe mechanical shock
can alter their operating characteristics or change the output levels, which
may require that the transducers be replaced.
P
T
REPARE TO
EST
Once you have performed the preliminary checks and any routine maintenance checks
prepare to test patients.
Turn on the instrument and allow it to come to operating temperature
(approximately 10 minutes).
Check that the transducers and other system components are operating properly.
Seat the patient comfortably in the test area.
Place the selected transducers on the patient.
Do not connect or disconnect Earphones, Insert Phones, Bone Vibrator, High
Frequency transducers, or any other accessories while in contact with the
patient.
D-0131088 Rev C, 2022-07
GSI AudioStar Pro Instructions for Use
A
S
P
UDIO
TAR
RO
Page 24 of 84