ONFIGURATION AND
The configure button presents a Configuration screen where it is
possible to view the instrument information such as serial number,
software version and last calibration date. This screen displays setup options to update the
AudioStar Pro software, configure bone conduction symbol settings, enable/disable the
Aux Intercom, export log files, set the date and time, and adjust the screen brightness.
Update
Place a USB drive with the appropriate update loaded into one of the four USB ports. Select
Update and then select from device, settings, or sound files to update the instrument.
Software and sound files updates must be obtained from GSI or an authorized GSI
representative. Selecting Settings allows for installing an update from a config file
generated from the Configuration Application.
Bone
Select the symbol scheme for bone conduction testing. Choose between MAS (Mastoid) and
FOR (Forehead.) This selection will be active throughout the current session. When a new
session is started, the symbol scheme will revert to the configured preference.
D-0131088 Rev C, 2022-07
GSI AudioStar Pro Instructions for Use