INSTRUCTIONS FOR USE
WARNING:
Before you remove and replace electrical components (power cord, pendant,
fuses), make sure to remove power from the Allen Advance Table. Failure to do
so could cause patient fall, injury, or equipment damage.
WARNING:
Do not service the unit while a patient is on the unit. Consult with facility
caregivers to make sure that a patient is not on the unit before you service the
unit. Patient injury can occur.
The parts shown below can be replaced by facility-authorized maintenance persons. If
additional parts need to be replaced, contact Allen Medical Technical Support.
Pendant
Pendant nut
H-bracket
Cross pin
Power cord, EU
Power cord, UK
Power cord, US
Power cord, IND/SA
Power cord, AUS
Power cord, China
Fuse, 5 mm x 20 mm, 6.3 A, 250 V, slow blow
(A-71100-XX*)
Fuse, 5 mm x 20 mm, 3.15 A, 250 V, slow blow
(A-71101-XX*)
* XX is a placeholder for the country code.
Document Number: 80028190
Version: E
Description
Quantity
1
1
1
1
1
1
1
1
1
1
2
2
Page 58
English
Part Number
A-71403
C-007239
A-71402
A-71401
A-71407
A-71406
A-71405
A-71408
A-71409
A-71412
C-007318
C-007358
Issue Date: 16 MAY 2022
Ref Blank Template: 80025117 Ver. F