INSTRUCTIONS FOR USE
Guidance and Manufacturer's Declaration — Electromagnetic Immunity
The A-71100-XX/A-71101-XX are intended for use in the electromagnetic environment
specified below. The customer or the user of the A-71100-XX/A-71101-XX should make
sure it is used in such an environment.
IEC 60601
Immunity Test
Test Level
Power
3 A/m
Frequency
(50/60Hz)
Magnetic Field
IEC 61000-4-8
NOTE: U
is the AC mains voltage prior to application of the test level.
T
Document Number: 80028190
Version: E
Compliance
Electromagnetic Environment—
Level
Guidance
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Page 16
English
Issue Date: 16 MAY 2022
Ref Blank Template: 80025117 Ver. F