Télécharger Imprimer la page

Hillrom Allen Advance Table Mode D'emploi page 17

Publicité

INSTRUCTIONS FOR USE
Guidance and Manufacturer's Declaration — Electromagnetic Immunity
The A-71100-XX/A-71101-XX are intended for use in the electromagnetic environment
specified below. The customer or the user of the A-71100-XX/A-71101-XX should make
sure it is used in such an environment.
IEC 60601
Immunity Test
Test Level
3 Vrms
Conducted RF
150 kHz to
IEC 61000-4-6
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher the frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
Document Number: 80028190
Version: E
Compliance
Electromagnetic Environment—
Level
Guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the A-71100-XX/A-71101-XX,
including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
d = 1.17√ P
d = 1.17√ P
3 Vrms
d = 2.33√ P
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
3 V/m
manufacturer and d is the
recommended separation in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey
compliance level in each frequency
range
vicinity of equipment marked with the
following symbol.
Page 17
English
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
a
, should be less than the
b
. Interference may occur in the
Issue Date: 16 MAY 2022
Ref Blank Template: 80025117 Ver. F

Publicité

loading

Ce manuel est également adapté pour:

A-71100-sch