Treatment; Cleaning And Sterilization; Preliminary Recommendations - Boston Scientific Swiss LithoClast Select System Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TREATMENT

In order to ensure contact between the stone and the pneumatic
probe, the probe extension must be carefully adjusted so that the end
of the pneumatic probe is flush with the end of the endoscope. There
is a risk of mucosal perforation if the pneumatic probe extends beyond
the endoscope.
Adjusting the position of the probe tip
Also make sure that the tip of the Pneumatic Probes does not
extend beyond the tip of the endoscope by more than 10 to 20 mm
when the probe is fully inserted into the working channel of the
endoscope.
10
FB-412_US ed 2008-04 User man pn3 BSC.indd 10

CLEANING AND STERILIZATION

Preliminary recommendations

Please conform to the recommendations of the "Reprocessing
Instructions" manual delivered with the product regarding procedure
of cleaning, disinfecting, packaging and sterilizing of the components.
Please follow the applicable present day sterilization regulations
in your country.
Do not disconnect the silicone air supply tube from the Pneumatic
Handpiece for cleaning, disinfection and sterilization. Handpiece
damage will occur if moisture enters the handpiece interior.
6/3/2015 2:35:30 PM

Publicité

Table des Matières
loading

Table des Matières