• Fully immerse attachment in cleaning solution in a suitable
container and manually agitate for 2 minutes.
• Use a sponge, soft lint-free cloth or soft bristle brush to
remove all visible soil.
• Gently use a brush to clean cannulations of attachment while
immersed in cleaning solution. Use ACB for all attachments
except MA-D20-G1 and Minimal Access Attachment Bearing
Sleeves. Use SDCB for MA-D20-G1 and Minimal Access
Attachment Bearing Sleeves. If attachment has a release
sleeve (Figure 1), move sleeve and brush exposed area.
Note: Brushes may be inserted from distal or proximal end of
attachments and bearing sleeves.
Warning: ACB and SDCB are single use and should only be used
to clean one set of attachments.
Caution: For MA-D20-G1 and Minimal Access Attachment Bearing
Sleeves use SDCB only. Do not force ACB or SDCB into or through
attachments or bearing sleeves. Do not use pipe cleaners.
Figure 1
4. Rinse Thoroughly
• Rinse device thoroughly using running hot deionized, distilled,
or purified water for a minimum of 2 minutes.
• Use a syringe or pipette filled with hot deionized, distilled or
purified water to flush lumens or channels.
• Manipulate any moving parts such as release sleeves (Figure
1).
Caution: Do not use high pressure water or air. Do not rinse
attachment with saline solution.
5. Inspect
• Visually inspect attachment for visible soil. Repeat Steps 2
thru 5 if visible soil remains.
Release Sleeve
5