English
Manual Cleaning
Warnings:
• Cannulations must be cleaned with appropriate brushes as
described in cleaning instructions.
• At the point of use the device must be cleaned as soon as possible
after use to prevent drying of blood, tissue, other biological debris
and contaminants on the device.
• National regulations must be observed.
• You must follow standard hospital procedures.
• You must follow manufacturer's recommendations for any
detergents, disinfectants and washing equipment used.
• Do not reprocess dissection tools. Dissection tools are disposable
and SINGLE USE ONLY.
Cautions:
• Rx Only: United States Federal law restricts this device
to sale by or on order of a physician or other licensed healthcare
provider.
• Do not use any alkaline detergent with a pH greater than 11. Alkaline
detergents with higher pH attack grease and seals which can
increase wear and cause device to malfunction.
• Do not use ultrasonic equipment.
• Do not use hypochlorite solution (i.e. bleach).
• Handpieces must not be immersed in cleaning solution but may be
sprayed with it.
• Fully demineralized (deionized, distilled, or purified) water must be
used during final rinse to prevent corrosion and spots.
• The Minimal Access Attachment Bearing Sleeves are non-repairable.
Reuse until bearing wear is detected and then replace. No bearing
sleeve service or repair will be provided.
Note: Use of alkaline detergents can cause device color to fade but this
does not impair function.
Operating Room (OR) Personnel Instructions
1. After completion of procedure, remove external debris by wiping
attachments and handpiece.
2. Dispose of open or used dissection tools.
2