Caution: Do not use high pressure water or air. Do not rinse attachment
with saline solution.
5. Inspect
• Visually inspect attachment for soil. Repeat Steps 2 thru 5 if visible
soil remains.
6. Go to Mechanical/Automated Cleaning section for further instructions.
Warning: Manual pre-cleaning must be followed by mechanical/
automated cleaning.
Mechanical/Automated Cleaning
Note: Prior to using G1 Cleaning and Sterilization Basket for first time, attach
connector sleeve to basket by clipping onto internal support structure (Figure
2). The connector sleeve should remain with basket for future use.
1. Loading the Basket
• Load devices in G1 Cleaning and Sterilization Basket by securing
them in appropriate brackets, as shown in G1 Cleaning and
Sterilization Basket Loading Diagram, with distal end angled down
where applicable.
• Place connector sleeve on connector end of electric handpiece
(Figure 3).
Caution: Connector end of electric handpiece must be sealed with
connector sleeve provided with basket (Figure 3).
Warning: Consult G1 Cleaning and Sterilization Basket Loading Diagram
for validated loading configurations. Angled attachments must be loaded in
the correct orientation as shown in Figure 4.
Figure 2
Figure 4
2. Mechanical/Automated Cleaning Cycle Process Parameters
12
Figure 3