Cycle
Rinse
Pre-wash
Cleaning
Rinse
Thermal
Processing
Dry
Caution: Do not use neutralizing agent.
3. Inspect
• Visually inspect devices for soil. Repeat manual pre-clean
and mechanical/automated clean if visible soil remains.
• Remove connector sleeve from connector end of handpiece
and drain until no visible droplets are coming from the
handpiece and handpiece connector.
4. Lubricate
Caution: This step is for attachments only. Do not lubricate
handpieces.
Option 1: 05.001.078, Lubricant for Anspach Systems, 110 ml
Warning: Do not use Gravity Air Displacement Steam Autoclave
Sterilization Process.
• For the following attachments apply one (1) spray pump
spray to the proximal end of the attachment.
SHORT-G1
SHORT-HD-G1
MEDIUM-G1
MEDIUM-HD-G1
Duration
(minimum)
2 minutes
1 minute
2 minutes
5 minutes
5 minutes
40 minutes
LONG-G1
LONG-HD-G1
MIA16-G1
CRANI-A-G1
Cleaning Instructions
Cold deionized (DI) or
purified water (PURW)
Warm DI or PURW water
(≥ 40°C); use detergent
Warm DI or PURW water
(≥ 45°C); use detergent
Rinse with DI or PURW
water
Hot DI water, ≥ 93°C
≥ 90°C
CRANI-L-G1
CRANI-P-G1
MA-10S
MA-10C
MA-15S
MA-15C
MA-15ST
MA-19ST
13