Anspach G1 Mode D'emploi page 14

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 75
English
beyond last bristle. Do not force ACB into handpiece. Do not use pipe
cleaners.
4. Rinse Thoroughly
• With distal end pointing down rinse device thoroughly using running
hot deionized, distilled or purified water for a minimum of 2 minutes.
• Hold handpiece with distal end pointing down and flush inside of
distal end using a syringe or pipette filled with hot deionized, distilled
or purified water.
Caution: Do not use high pressure water or air. Do not immerse. Do not
allow running water to enter distal end of handpiece or console connector.
5. Inspect
• Visually inspect handpiece for visible soil. Repeat Steps 2 thru 5 if
visible soil remains.
6. Go to Mechanical/Automated Cleaning section for further instructions.
Warning: Manual pre-cleaning must be followed by mechanical/
automated cleaning.
Manual Pre-Cleaning of Attachments
Note: The instructions in this section are specific to the G1 Attachments
and MA Bearing Sleeves listed in Table 1.
1. Preparation
• Assemble all necessary supplies including Attachment Cleaning
Brush (ACB), Small Diameter Cleaning Brush (SDCB), sponge, lint-
free cloth and soft bristle brush.
• Remove attachment and dissection tool from handpiece.
2. Rinse
• Rinse attachment under running cold deionized, distilled or purified
water for a minimum of 2 minutes.
• Manipulate any moving parts such as release sleeves under running
cold deionized, distilled or purified water to loosen or remove gross
soil.
• Use a sponge, soft lint-free cloth or soft bristle brush to assist in
removing gross soil.
Caution: Do not use high pressure water or air. Do not rinse attachment
with saline solution.
10

Publicité

Table des Matières
loading

Table des Matières