• For the following attachments apply one (1) spray pump spray to the
distal end of the attachment and one (1) to the proximal end of the
attachment.
MA-D20-G1
CSR60-G1
• Applying excess amounts of lubricant will cause the lubricant to drip
from the attachment. Clean off any excess lubricant.
Option 2: Non-silicone based medical lubricant
• Prepare a lubricating solution of instrument milk (non-silicone based
medical lubricant) per the manufacturer's directions.
• Fully immerse attachment in lubricating solution, at room
temperature, in a suitable container and agitate for 15 seconds.
• Remove attachment and allow it to drain completely until no visible
droplets are coming from it.
• Return attachments to basket for further processing.
Caution: Do not rinse out instrument milk (lubricant). Do not apply mineral
oil or other lubricants, which may cause attachment to overheat.
5. Go to Sterilization section for further instructions.
Note:
• The clinical processing instructions provided have been validated
by DePuy Synthes Power Tools for preparing a non-sterile medical
device; this instruction is provided in accordance with ISO 17664
and ANSI/AAMI ST81.
Sterilization
Before sterilizing:
• Ensure sterilization equipment is in proper working order as
specified by manufacturer.
• Assure equipment manufacturer's instructions are properly
employed by trained and qualified personnel.
• Assure actual cycle employed has been properly validated for
the device(s)/load configuration being processed and appropriate
sterilization indicator devices are included for each process and
cycle.
14
QD8-G1
QD11-G1
QD8-S-G1
QD11-S-G1
QD14-G1
QD14-S-G1