Reprocessing Personnel Instructions
Handpiece
Note: The instructions in this section are specific to the eG1 Electric
Handpiece.
1. Preparation
• Assemble all necessary supplies including Attachment
Cleaning Brush (ACB), sponge, lint-free cloth and soft bristle
brush.
• Remove attachment and dissection tool from handpiece.
2. Rinse
• Rinse handpiece with distal end pointing down under running
cold deionized, distilled or purified water for a minimum of 2
minutes.
• Use a sponge, soft lint-free cloth or soft bristle brush to assist
in removing gross soil.
Caution: Do not immerse. Do not use high pressure water or air.
Do not allow running water to enter distal end of handpiece or
console connector (end of cable).
3. Clean
• Prepare cleaning solution using alkaline detergent or neutral
pH enzymatic detergent per manufacturer's directions for
correct temperature, water quality (i.e. pH, hardness) and
product concentration/dilution.
• With distal end pointing down spray and wipe using cleaning
solution for a minimum of 2 minutes.
• Use a sponge, soft lint-free cloth or soft bristle brush to
remove all visible soil.
• Hold handpiece with distal end pointing down and insert ACB
wetted with cleaning solution into distal end of handpiece.
Warning: ACB is single use and should only be used to clean one
handpiece and associated attachments.
Caution: Do not immerse. Do not allow running water to enter
distal end of handpiece or console connector. Do not insert ACB
beyond last bristle. Do not force ACB into handpiece. Do not use
pipe cleaners.