Note: Prior to using G1 Cleaning and Sterilization Basket for first
time, attach connector sleeve to basket by clipping onto internal
support structure (Figure 2). The connector sleeve should remain
with basket for future use.
Warning: Devices must be cleaned prior to sterilization using the
validated cleaning method outlined in these instructions.
1. Loading the Basket
• Load devices in G1 Cleaning and Sterilization Basket by
securing them in appropriate brackets, as shown in G1
Cleaning and Sterilization Basket Loading Diagram, with
distal end angled down where applicable.
• Remove connector sleeve from handpiece connector (Figure 1).
Replace connector sleeve in its bracket and lay handpiece
connector in bottom of basket.
Warning: Handpiece connector sleeve must be removed from the
handpiece connector prior to sterilization (Figure 1). Consult G1
Cleaning and Sterilization Basket Loading Diagram for validated loading
configurations. Angled attachments must be loaded in the correct
orientation as shown in Figure 2.
Figure 1
2. Packaging
• When wrapping the basket the lid must be used.
• In US double wrap basket or instruments with an FDA cleared
sterilization wrap using appropriate wrapping techniques
described in ANSI/AAMI ST79 (For example: Simultaneous
double wrapping or sequential wrapping).
• Outside US double wrap basket or instruments in accordance
with local procedures, using standard wrap and wrapping
techniques such as those describe in ANSI/AAMI ST79.
Warning: When using a sterilization wrap the basket lid must be
used. Must follow manufacturer's instructions for any wraps used.
Figure 2