4. Rinse Thoroughly
• With distal end pointing down rinse device thoroughly using running
hot deionized, distilled or purified water for a minimum of 2 minutes.
• Hold handpiece with distal end pointing down and flush inside of
distal end using a syringe or pipette filled with hot deionized, distilled
or purified water.
Caution: Do not immerse. Do not use high pressure water or air. Do not
allow running water to enter distal end of handpiece or console connector
(end of cable).
5. Inspect
• Visually inspect handpiece. Repeat Steps 2 thru 5 if visible soil
remains.
6. Go to Sterilization section for further instructions.
Attachments
Note: The instructions in this section are specific to the G1 Attachments
and MA Bearing Sleeves listed in Table 1 in the Mechanical/Automated
Cleaning section of this document.
1. Preparation
• Assemble all necessary supplies including Attachment Cleaning
Brush (ACB), Small Diameter Cleaning Brush (SDCB), sponge, lint-
free cloth and soft bristle brush.
• Remove attachment and dissection tool from handpiece.
2. Rinse
• Rinse attachment under running cold deionized, distilled or purified
water for a minimum of 2 minutes.
• Manipulate any moving parts such as release sleeves under running
cold deionized, distilled or purified water to loosen or remove gross
soil.
• Use a sponge, soft lint-free cloth or soft bristle brush to assist in
removing gross soil.
Caution: Do not use high pressure water or air. Do not rinse attachment
with saline solution.
3. Clean
• Prepare cleaning solution using alkaline detergent or neutral pH
enzymatic detergent per manufacturer's directions for correct
temperature, water quality (i.e. pH, hardness) and product
concentration/dilution.
4