Technical inspections
For safety, the devices require regular
maintenance. To maintain the functional
and operational safety, check all compo-
nents for damage and loose connections at
least once a year.
These checks should be performed by
persons with adequate training and expe-
rience. Damaged and worn parts must im-
mediately be replaced with original spare
parts by authorized sta .
Warning!
Patient hazard, equipment malfunction
and damage –
– Refer repair and maintenance to
authorized persons. Persons are
authorized after training by a special-
ist trained and commissioned by the
manufacturer.
ORMED GmbH will make all docu-
ments required for servicing, such as
circuit diagrams, parts lists, descrip-
tions or calibration instructions,
available to authorized experts.
The inspections can be carried out by DJO
Global Technical Service within the frame-
work of a service agreement.
The device does not require additional
regular maintenance.
Fuse replacement
Warning!
Patient hazard, equipment
malfunction and damage –
The replacement of fuses must be
referred to specialists as defined in IEC
60364 or other applicable standards
(e.g. biomedical technicians, electri-
cians, electronics installers).
Fuses used must be T A H250Vac fuses.
Before replacing fuses, turn o the
ARTROMOT®-SP and disconnect the
device from the power line.
Use an appropriate tool to remove the fuse
holder situated between the power switch
and the power connector (Fig. 1). Replace
the fuses and reinsert the fuse holder (Fig.
2). Ensure that the fuse holder properly
locks into place.
Fig. 1
Fig. 2
. . Transport
The following operating steps must
be completed before transporting the
ARTROMOT®-SP :
1.
Activate the "transport setting"
function
in the menu (see also
5.3) and start the ARTROMOT®-SP .
2. Remove the footplate (7) as well as its
bracket (11). To do so, simply pull it from
the square-end guide at motor B. (Fig.
G)
3. For transport of the device, motor B
must be mounted on the right side
of the moving bracket (15) (marked
R). Refer to section 6.4 Conversion, if
required.
4. Turn o the ARTROMOT®-SP power
switch.
5. Disconnect the power cord and the
programming unit.
6. Relieve the strain on the carriage by
lifting it a little at the moving bracket
for dorsal extension / plantar flexion
(15) and open the clamping lever which
adjusts the inclination of the carriage
(5). Select an angle where the carriage
is parallel with the adjusted height.
Tighten the clamping lever to lock the
carriage in this position.
.
Again lift the moving bracket for dorsal
extension / plantar flexion (15) a little
to remove the strain. Then loosen the
locking device for the height adjust-
ment by simultaneously pressing the
two black pins of the locking device
together (19). Select a height where
the locking pins engage in the top hole
(marked: Transport)
8. Only use the original shipping box for
transporting the device. ORMED GmbH
cannot be held liable for transport
damage, if the device was not shipped
in its original shipping box.
9. Pack the programming unit (22) in the
supplied box.
10. Place the footplate bracket (11) with the
footplate (7) and the power cord in the
extra box provided.
11. Place the ARTROMOT®-SP at the bot-
tom of the shipping box. Then pack the
two extra boxes for the programming
unit and the footplate + bracket in the
shipping box along with the padding
material, as shown below.
Extra box with
Extra box with
programming unit
footplate/
bracket and
power cord