Sirona C1+ Notice D'utilisation page 133

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Sirona Dental Systems GmbH
40000
#
24000
#
59 06 446 D 3386
D 3386.201.01.08.02
C
E1
C
E1
12.1 Instrument dialog: Electric motor
Changeover switch for operation as
direct starter or speed controller foot switch.
Unit lamp OFF / ON
If this key is pressed longer, the
The brightness setting dialog appears.
(see page page 24).
Motor speed range:
200 to 40000rpm.
The motor is started by actuating the foot switch.
Operation as direct starter (key appears with a normal
frame):
When actuating the foot switch, you will be working with
the fixed value programmed for the motor.
Operation as speed controller (key appears with a
bold frame):
An expanding triangle appears. It indicates the regula-
tion of the speed during motor operation.
The foot switch operates as a speed controller within a
range of 0 to the programmed speed.
Programming of functions
For each motor, different functions can be programmed
for any user (A, B or C; see Neutral dialog 0 or assistant
dialog, page 31 and 32) in the three function levels E1,
E2 and E3:
CAUTION
Prior to any change of user A, B or C, all instruments
must be placed back in their receptacles!
Remove motor from its receptacle.
The electric motor instrument dialog appears.
Select the desired function level E1, E2 or E3.
Set the desired maximum motor speed.
Select the desired functions by actuating the corre-
sponding keys.
Example shown for user C:
(symbols with bold frame)
Coolant
Direction of rotation,
clockwise
max. motor speed
24000 rpm
Light on the motor
12 Instrument dialogs
Function level E1
controller foot switch
37

Publicité

Table des Matières
loading

Table des Matières