2
4
6
12:45
11 Programming automatic positioning
Run the activated program
until the end position is reached*!
The selected program symbol appears with a bold frame
now (example: program 1).
•
Switch the unit lamp ON or OFF. This state is then al-
so programmed.
•
Actuate the foot switch.
The cursor • then jumps to the symbol for the manual
chair setting dialog.
•
Activate the
manual chair setting dialog.
The program appears.
•
Actuate the different setting keys to move the chair
into the desired treatment position ( see page page
29).
•
Then bring the dentist element into the desired posi-
tion by hand.
Observe the safety information "Chair and
dentist element" on page 9!
Place the cursor • on the programming symbol (if
•
programming with the foot switch).
•
Actuate the programming key until three successive
acoustic signals sound.
Programming is completed now.
i
NOTE
* If the program has not run up to the end position prior to
resetting it (i.e. if it is aborted), the new treatment position
cannot be programmed! This is indicated by the dimmed
programming symbol.
33