21 Additional equipment and accessories

21.1 Media block on the patient chair

21
Additional equipment and
accessories
21.1
Media block on the patient chair
Technical data and requirements
82
Water
Air
Fuse
Line voltage
230 V
Additional equipment
Only approved additional devices (e.g. supplied by
EMS) may be connected to a media block on the base of
the lower part of the chair (the manufacturer's instruc-
tions must be complied with).
The media block is equipped with quick couplings for air
and water and an outlet for non-heating appliances with
a separate fuse (6.3A).
For treatment centers with a disinfection unit, the inte-
gration of a media block complies with the DVGW
approval (EN1717/ DIN1988). In this case, the addi-
tional devices connected are also compliant with the
requirements of EN1717 / DIN1988.
All connected devices must be supplied with disin-
fected water if a disinfection unit is installed.
0.1‰ -0.2‰ hydrogen peroxide are required for reg-
ular operation and 1.4% for sanitization.
CAUTION
Before connecting any additional devices, first check
whether they are suitable for hydrogen peroxide.
If these devices are not suitable for the higher concentra-
tion used for sanitization, Sirona recommends separat-
ing them from the treatment center during sanitization
and then sanitizing them separately according to the
manufacturer's instructions.
Sirona shall not be held liable for any damage to con-
nected devices!
Pressure and flow rate specifications
Pressure
2,2 ± 0.2 bar
Water
4,4 ± 0.5 bar
Air
Sirona Dental Systems GmbH
Flow rate
max. 300 ml/min
max. 70 Nl /min
59 06 446 D 3386
D 3386.201.01.08.02