22 Maintenance
22
Maintenance
22.1
Care and cleaning by the practice team
22.2
Inspection and maintenance
22.3
Safety tests
86
Despite the outstanding quality of your treatment center,
it is necessary to perform inspections and preventive
maintenance as well as safety tests at predetermined
intervals in order to ensure its operational safety.
These events are documented in the Maintenance
Manual.
Care and cleaning is done by the practice team.
For details please refer to the following sections.
The practice team is responsible for regular cleaning
and care of the treatment center, with the aim of
minimizing the risk of contamination for patients and
users.
The work to be performed is specified in the document
"Care and Cleaning by the Practice Team".
In order to guarantee the operational safety and
reliability of your treatment center and to avoid damage
due to natural wear, yearly inspection and maintenance
must be performed on your treatment center. This is
done by an authorized service engineer of your dental
depot.
The steps to be performed as well as the parts which
must be replaced are specified in the document
"Maintenance Certificate".
An overview of the inspection and maintenance work
performed is additionally entered by the service
engineer in the "Installation Report / Warranty
Passport". This document is part of the "Maintenance
Manual".
Medical products are designed in such a way that the
first occurrence of a fault does not create a hazard to the
safety of patients, users or other persons. Hence it is
important to detect such faults before a second fault
occurs, which might then lead to safety hazards.
For that reason it is essential to perform safety tests
every 2 years which aim particularly at detecting
electrical faults (e.g. isolation defects). This is done by
an authorized service engineer from your dental depot,
most practically together with the work to be performed
according to section 22.2.