4 OPERATION
4.4 Test Menu: Press
The following compensations and controls are carried out in the Test menu:
OFFSET
SPAN
0......9
T° ADJUST
SIMUL
END

4.4.1 Offset-Compensation

In order to check and modify the basic setting of 4 mA, connect an ammeter in the output
circuit. Press ENTER when "OFFSET" is displayed, the transmitter generates 4 mA. If the
measured value is different from 4 mA, enter the measured value as offset value.
OFFSET
ENTER
SPAN
ENTER

4.4.2 Span-Compensation

Check and modify the basic setting of 20 mA. The procedure is identical to the Offset-com-
pensation. The transmitter generates 20 mA, if the ENTER key is pressed when "OFFSET"
is displayed. Correct the span value by entering the measured value if necessary.
SPAN
ENTER
T ° ADJUST
ENTER
ENTER
simultaneously for 5 seconds
0......9
Zero point compensation (4 mA) (§ 4.4.1)
Span compensation (20 mA) (§ 4.4.2)
Temperature value correction. The outputs reacts according to
this input (§ 4.4.3)
pH or temperature value to be simulated. The outputs react
according to this input (§ 4.4.4)
Return to the operation mode and storage of new parameters
for OFFSET and SPAN and T° adjustment.
If one of the values is erroneous, the transmitter displays "OFF-
SET", and new values must be entered.
OF= 04.00
0......9
OF= 04.02
Enter the measured value
SP= 20.00
0......9
SP= 19.96
Enter the measured value
E-32-
PH TRANSMITTER 8205