Safety instructions
10
DANGER
If the voltage measuring lead is not used as intended, the
protection provided by the voltage measuring lead may
be impaired.
DANGER
The ITC-U1001 is intended solely for use on a Quan
tumX MX809B. Failure to comply with the documentation
for the MX809B may cause the protection provided by
the voltage measuring lead to be impaired.
DANGER
If the ITC-U1001 voltage measuring lead is used outside
the ratings, the protection provided by the voltage mea
suring lead may be impaired.
DANGER
Outside of the measurement categories acc. to EN
61010 (not connected to the supply network, formerly:
CAT I) the ITC-U1001 is only approved for a maximum
voltage to ground potential of 1000 V DC or 1000 V AC
rms if the following requirements are met:
S An analysis has been performed of the working
voltage, loop impedance, occasional overvoltage and
transient overvoltages of the (potential) circuit to be
contacted.
S The results of the analysis fall within the allowable
values according to the data sheet.
A4749-1.0 HBM: public
ITC-U1001