Thermocouple
Transducers that can be con­
nected
Permissible line lengths between
MX809B and transducer, max.
Linearization ranges
Type B (Pt‐30 % Rh and Pt-6 % Rh)
Type C (W and W-26 % Re)
Type E (Ni-Cr and Cu-Ni)
Type J (Fe and Cu-Ni)
Type K (Ni-Cr and Ni-Al)
Type N (Ni-14.2 % Cr and Ni-4,4 %
Si-0.1 % Mg)
Type R (Pt-13 % Rh and Pt)
Type S (Pt-10 % Rh and Pt)
Type T (Cu and Cu-Ni)
Transducer impedance
Signal bandwidth (-3 dB)
Noise type K (peak-to-peak)
with 1 Hz Bessel filter
Total error limit at 22
temperature
Types E, J, K, N, T, C
Types R, S
Type B
Temperature drift (type K)
Optionel post‐scaling of the tem­
perature values
Number of pairs of values in the
MX1609KB, max.
Number of pairs of values from
TEDS, max. (from Template
Calibration Table)
MX809B
C [F]
C [F]
C [F]
C [F]
C [F]
C [F]
C [F]
C [F]
C [F]
o
C ambient
K/10 K
A3757-2.2 HBM: public
(types B, C, E, J, K, N, R, S, T)
m
+100 ... +1820 [212 ... 3308]
0 ... +2300 [32 ... 4172]
-200 ... +900 [−328 ... 1652]
-200 ... +1200 [−328 ... 2192]
-100 ... +1300 [−148 ... 2372]
-270 ... +1300 [−454 ... 2372]
-50 ... +1768 [−58 ... 3214]
-50 ... +1768 [−58 ... 3214]
-270 ... +400 [−454 ... 752]
Hz
K
K
K
K
Specifications
Thermocouples
30
< 500
55
0.2
± 1
± 4
± 15
<"0.4
64
14
53