NOTE!
Automatic scale detection
The pump's software automatically detects whether the container scale and bag
scale are connected or only the container scale.
NOTE!
Deficit accuracy
The greater the consumption of irrigation fluid, the greater the deviation between
the actual and the displayed deficit (see Chapter Technical Data [} 49]).
Try to collect all the fluid running out of the uterine cavity during the procedure to
achieve the most exact deficit value possible.
The container scale can be loaded with a weight of up to 25 kg (55 lbs). The max-
imum load of the bag scale is 12 kg (27 lbs). If the scales are loaded beyond these
limits, the message Scale Overload/Check Scale is triggered and displayed. Three
audible warning beeps are emitted as well (see Chapter Error and Warning Mes-
sages [} 48]).
WARNING!
Containers and fluid bags
Make sure the containers and fluid bags hang freely, are not resting on something,
and do not touch other objects except the bag deflectors. Failure to follow these in-
structions means the deficit cannot be calculated correctly.
NOTE!
Scale recognition
Connect the scale to the pump before turning the system on to ensure the system
recognizes the scale.
System Operation
Precise balancing
Scale capacity