CAUTION!
Modifying the system
Modifying the system and/or its accessories is not permitted.
CAUTION!
Combination of low set pressures and excessive vacuum pressures
When using the Aquilex® Fluid Control System with tissue removal systems, e.g.
MyoSure®, the combination of low set pressures and excessive vacuum pressures
may result in a significant loss of intrauterine distension pressure which has the
potential to affect the visibility of the surgical field. Conversely, when employing a
high distension pressure, the deactivation of the tissue removal system can lead to
pressure spikes that can exceed 150 mmHg.
CAUTION!
Do not use the covered power output at the rear of the irrigation pump unit.
CAUTION!
Hysteroscope
The system may only be connected with hysteroscopes designed for and featuring
the technical specification permitting such a combined use. Any utilized hystero-
scopes must comply with the most recent versions of IEC 60601-2-18 and ISO 8600.
CAUTION!
Electrical interference
(see Chapter Electromagnetic Compatibility [}  51]): Electrical interference with
other devices or instruments was considered when developing this system and
none was detected during testing. However, if you still detect or suspect such inter-
ference, please follow these suggestions:
• Move the Aquilex® System, the other device, or both devices to a different loc-
ation.
• Increase distance between devices used.
• Consult an electro-medical expert.
CAUTION!
Wall outlet voltage
Check to ensure the available wall outlet voltage matches the data listed on the la-
bel attached to the back of the pump. Incorrect voltage can cause errors and mal-
functions and may destroy the system.
CAUTION!
Transport
The device is transportable. The roller wheels of the Fluid Monitoring Unit (cart/
scale) are used for positioning at the place of use. To transport the device, remove
all fluid bags from the hooks and make sure there are no containers or only com-
pletely emptied containers on the cart/scale. Inflow and outflow tubes must be
completely removed. Make sure the power supply line does not touch the ground
and there are no other objects located on the Aquilex® Fluid Control System. Al-
ways use the handle to move the system safely.