5.8

Connecting the Outflow Tube Set

CAUTION!
Combination of low set pressures and excessive vacuum pressures
When using the Aquilex® Fluid Control System with tissue removal systems, e.g.
MyoSure®, the combination of low set pressures and excessive vacuum pressures
may result in a significant loss of intrauterine distension pressure which has the
potential to affect the visibility of the surgical field. Conversely, when employing a
high distension pressure, the deactivation of the tissue removal system can lead to
pressure spikes that can exceed 150 mmHg.
(3)
(1)
(2)
(7)
(1)
Using low vacuum configuration of Fig. Outflow tube set [}  31], connect outflow
tube set (Y-tube) to patient port (4) of second container. Yellow flexible connector
attaches to drape (6). Yellow Luer fitting connects to stopcock (7) of removable
outflow channel or hysteroscope outflow valve.
(4)
(5)
(6)
(3)
Fig. 5–9  Outflow tube set
(1)
To low vacuum port (white)
(2) Container
(3) Connecting tube
(4) Patient port
(5) Outflow tube set
(6) Drape
(7) Removable outflow channel or
hysteroscope outflow sheath
stopock
(5)
System Operation
en
31 / 378