System Operation
Open outer packaging
en
Connect to hysteroscope
Inserting the tube set
Connect the fluid bags
Fig. 5–13  Attach roller wheel tube
(5) Inflow tube
(6) Roller wheel tube
(7)
Pressure chamber
(8) Hyteroscope tube
(11) Roller wheel
(12) Holder for inflow tube
(13) Alignment arrows
Intrauterine pressure setting
Safety threshold
34 / 378
• Inflow tube set - To be carried out by non-sterile technician:
– Open outer packaging of the inflow tube set.
– A sterile nurse then removes the inner tube set package and opens it.
• To be carried out by sterile technician:
– Keep the blue Luer lock connector (9) in the sterile area and hand the
tube end with the bag spikes (2) to the non-sterile technician.
– Connect the blue Luer lock connector (9) with the hysteroscope inflow
stopcock. Open stopcock.
• To be carried out by non-sterile technician:
– Ensure system is turned on.
– Close the clamps (3) on the inflow tube below the bag spikes (2).
– Insert the inflow tube set into the inflow tube holder. Insertion of the
roller wheel tube is depicted in Fig. Attach roller wheel tube [} 34].
– Carefully insert the pressure chamber (7) into the lower notch of the in-
flow tube holder (12) until you feel resistance. Align pressure chamber and
inflow tube holder using arrows (see Fig. Attach roller wheel tube [} 34]).
– When inserting the roller wheel tube, make sure not to damage the mem-
branes of the pressure chamber. Insert the pressure chamber (7) only if
chamber is not pressurized.
– Place the roller wheel tube (6) around the roller wheel (11).
• When connecting or removing the tube to or from the irrigation fluid bags, al-
ways grasp the bag spike at the provided handle. Observe aseptic technique
when inserting the spike(s) into the fluid bag(s). The surgeon must select a dis-
tension fluid suitable for the type of procedure.
(5)
(12)
(8)
(7)

5.10 Presetting the Intrauterine Pressure

The intrauterine pressure setting can be adjusted while the system is in operation.
Use the º and » buttons (Fig. Front of irrigation pump unit [}  22]). The pressure
setting can be adjusted to between 40 to 150 mmHg in steps of 5 mmHg.
The intrauterine pressure is shown on the intrauterine pressure display (2).
If when scrolling with the º button (Fig. Front of irrigation pump unit [}  22]) the
safety threshold of 100 mmHg is reached, an audible warning signal is emitted. Re-
lease the º button for one second and scroll again to set higher values up to
150 mmHg.
(6)
(11)
(13)