4
Initial System Set-Up
Always check all parts and accessories of the system when performing initial setup.
If the system has obvious defects, contact Hologic Technical Support (Chapter War-
ranty Information [} 55]).
Place the system on a level surface and install in a dry environment. The ambient
temperature and humidity must meet the requirements mentioned in Chapter
Technical Data [} 49].
CAUTION!
Device setup
Device should be positioned outside of the sterile area in such a way that
• it can be easily disconnected,
• it is easy to use and switch off and on,
• it allows an easy monitoring of the display values, device functions, and access
to the control elements.
CAUTION!
Indoor climate
Before switching on the device, sufficient time must have passed to adjust to the
indoor climate.
4.1
Preparing the System for Use
CAUTION!
Possible malfunctions
The device Aquilex® Fluid Control System should not be used directly next to other
devices as this could result in malfunctions. The device Aquilex® Fluid Control Sys-
tem was tested for compliance with IEC 60601-1-2 as a standalone system. There-
fore, do not stack other devices (e.g. MyoSure® Control Unit) on the system or the
Irrigation Pump Unit. In particular, do not place any other device than the
AQL-100PBS on the trays of the AQL-100CBS. If usage in the manner described
above is nevertheless required, this system and the other devices should be mon-
itored to make sure they function properly.
CAUTION!
Combination of AQL-100PBS with AQL-100CBS
The irrigation pump unit AQL-100PBS may only be used with the fluid monitoring
unit AQL-100CBS as only this combination is approved to bear the MET NRTL mark.
CAUTION!
Mains connection
• Make sure the available mains voltage matches the data listed on the type la-
bel attached to the back of the device. Incorrect voltage can cause errors and
malfunctions and may destroy the device.
• Make sure the connection data and technical specifications of the power sup-
ply comply with DIN VDE or national requirements. The mains connection
cable may be plugged only into a properly installed, grounded safety wall
socket (shockproof socket) (see DIN VDE 0100-710).
• Read the device label located in rear of device (type plate) to determine the op-
erating voltage of the device.
Initial System Set-Up
Initial system set-up
Mains connection
en
19 / 378