Purpose
en
14 / 378
WARNING!
Function test
The function test must be performed prior to each device use.
WARNING!
Obvious defects
Never use the system if it has suspected or confirmed defects, especially if these in-
volve the power plugs or the mains power supply connection cables. In this case
have the device repaired by authorized service personnel.
WARNING!
System error
Do not use the Aquilex® System if a defect is suspected or detected during the
function test. This also applies to any obvious defects, especially defects on the
power connector or plug and power cord.
WARNING!
Disconnect the power cord
Pressing the ON/OFF switch does not disconnect the system from the wall power
outlet. This requires pulling the power cord located in the rear of the system.
WARNING!
Technique and procedures
Only the physician can evaluate the clinical factors involved with each patient and
determine if the use of this device is indicated. The physician must determine the
specific technique and procedure that will accomplish the desired clinical effect.
WARNING!
Check all factory settings
Factory settings are not mandatory settings for the physician. The physician is re-
sponsible for all settings affecting the surgical procedure.
WARNING!
Original accessories
For your own safety and that of your patient, use only Aquilex® accessories (see
Chapter Accessory List [} 54]).
WARNING!
Additional equipment
Additional equipment connected to medical electrical devices must be demon-
strated to be compliant with their respective IEC or ISO standards (IEC  60601-1,
IEC 60950 or IEC 62368 for data processing equipment). Furthermore, all configura-
tions must comply with the normative requirements for medical systems (see sec-
tion 16 of the last valid edition of IEC  60601-1). Anyone who connects additional
devices to medical electrical equipment is a system configurator and as such is re-
sponsible for the system's compliance with the normative requirements for sys-
tems. Please contact the technical service if you have additional questions.