Setting Display Contrast (monochrome display)
1. Press the ▼ or ▲ button to adjust the contrast level.
2. Press the
Setting Backlight
1. Press the
2. Change option by pressing the ▼ or ▲ button.
3. Press the
4. Change timeout by pressing the ▼ or ▲ button.
5. Press the
Back To Main Menu
There are three options at this point; press the:
▼ button to go to the Sensor Options menu
▲ button to go to the PREVIOUS SETUP page in the Instrument
Options menu
button to go to the Instrument Options menu.

3.5. MSA Link Operation

Connecting Instrument to PC
1. Switch ON the ALTAIR 5X Gas Monitor and align the Datalink
Communication port on the ALTAIR 5X Monitor to the PC
IR interface.
2. Start the MSA Link software on the PC and start the connection by
clicking the connect icon.
See MSA Link documentation for detailed instructions.

3.6. Function Tests on the Instrument

Alarm Test
1. Turn ON the instrument. The user should verify that:
• alarm LEDs flash
• horn sounds briefly
• vibrating alarm triggers briefly.
button to confirm the contrast level.
button to enter Setup.
button to enter.
button to confirm timeout.
3-15