Setting Backlight
1. Press the
2. Change option by pressing the ▼ or ▲ button.
3. Press the
4. Change timeout by pressing the ▼ or ▲ button.
5. Press the
Back To Main Menu
There are three options at this point:
•
the ▼ button
•
the ▲ button
•
the
button
3.5. MSA Link Operation
Connecting Instrument to PC
1. Switch ON the ALTAIR 5 instrument and align the Datalink
Communication port on the ALTAIR 5 to the IR interface of the PC.
2. Start the MSA Link software on the PC and start the connection by
clicking the connect icon.
See MSA Link documentation for detailed instructions.
3.6. Function Tests on the Instrument
Alarm Test
1. Turn ON the instrument. The user should verify that:
• alarm LEDs flash
• horn sounds briefly
• vibrating alarm triggers briefly.
3.7. Calibration Check
The calibration check is simple and should only take about one minute.
Perform this calibration check before each day's use for each
installed sensor.
3-14
button to enter Setup.
button to enter.
button to confirm timeout.
to go to the Sensor Options menu
–
to go to the Previous Setup page in the
–
Instrument Options menu
to go to the Instrument Options menu.
–