Turning On - MSA Altair 5X Mode D'emploi

Masquer les pouces Voir aussi pour Altair 5X:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3.2 Turning ON

Instrument operation is dialog driven from the display with the aid of the
three Function buttons (see Section 2.2.1). For more information, see
the flow charts in Section 7.
Turn the instrument ON with the
The instrument performs self tests. During the self test, the
instrument checks alarm LEDs, audible alarm, vibrating alarm, and
installed sensors.
The instrument displays:
Startup logo
Software version, instrument serial number, company name,
department, and user names
Sampling system safety test; see Section 3.2.1
During the turn-ON sequence, if a sensor was changed since
the previous instrument operation, the current listing of the
installed sensors displays and user interaction is required.
• The user must accept the new configuration by pressing
the ▲ button.
• If the current sensor configuration is not accepted, the
instrument alarms and is not usable.
Combustible gas type, and installed sensor indication
Combustible gas type and sensor units
(monochrome display only)
Low Alarm setpoints
High Alarm setpoints
STEL Alarm setpoints (if enabled)
TWA Alarm setpoints (if enabled)
Calibration cylinder settings
Current date
Last calibration date (if enabled)
Calibration due date. If the calibration due date is enabled,
the message "CAL DUE; X DAYS" appears on the
instrument display.
X = the number of days until a calibration is due,
3-2
user selectable for 1 to 180 days.
button.

Publicité

Table des Matières

Dépannage

loading

Ce manuel est également adapté pour:

Altair 5x ir

Table des Matières