PNEUMATIC FOWLER RELEASE HANDLE ADJUSTMENT

English
Tools Required:
• #2 Phillips Screwdriver
Procedure:
1 .
Apply the brakes .
2 .
Using a #2 Phillips screwdriver, loosen the two screws on the handle linkage . Release the handle tension spring
on the patient left side .
3 .
Adjust the handles (B), so there is a 3/4" gap between the inner fowler frame and the handle .
4 .
Tighten the two screws on the handle linkage bar and reattach the spring .
5 .
Check the adjustment of the pneumatic cylinders . See "Pneumatic Fowler Adjustment" on
6 .
Verify proper operation before returning the unit to service .