Electrolyte
Flow
Pencil and Cathode
Polishing Chamber
Power Supply
18
Movipol-3
Instruction Manual
2. Trouble-shooting
The Figures referred to in this manual can be found
in the Spare Parts and Diagrams section.
Polishing defects can normally be corrected using the following
trouble-shooting schedule:
Check the age of the mixed electrolyte. It should not be more
than 2-3 months old. The A-2 electrolyte has a particularly short
lifetime (approximately 2 months) in a mixed condition.
Check the number of polishings made with the electrolyte. It
should not be used for several hundred polishings.
Check that the right electrolyte - material combination is used.
The pump suction is adjusted to 300 "20 mmhg at the factory
(some materials, such as Nimonic and Titanium alloys need a
flow corresponding to 400-450 mmhg). The suction should be
measured with a vacuum meter using the suction hose nipple
(fig. 7.3).
Check that all connections are tight and that air is not sucked
into the stream of electrolyte.
The pump suction (flow) can be adjusted by means of the screw
(fig. 7.2). Screw inward increases the flow. Screw outward
decreases the flow.
Check that the cathode (fig. 9.100) is even with, or retracted
approximately 0.5 mm from the edge of the pencil tip without
polishing chamber.
Check that the preset voltage can be established between the
cathode and the specimen (the anode).
The resistance can be very high in an electrolyte which is
incorrectly mixed.
Check that the polishing chamber is properly mounted and still
flexible. If not flexible, exchange the chamber; see section
INSTALLATION.
Check that the batteries are charged.
Check that the transformer voltage is 42V "10%.
Note: