Mounting the Pump-Container
Unit
Dismounting the Pump-
Container Unit
Filling Electrolyte into the
Container
Emptying the Container of
Electrolyte
Mounting the Polishing
Chamber
8
Movipol-3
Instruction Manual
Lower the pump-container unit into the cabinet.
Press the unit down until a click is heard. The pump-container
unit is now locked in the cabinet.
Press the two black locks to the sides from the pump in direction
of the cabinet wall, a click is heard and the pump-container unit
is dismounted and can be removed from the cabinet.
Before filling electrolyte into the container, the section, Safety
Precautions should be carefully observed.
Remove the pump and container from the cabinet.
Remove the pump from the electrolyte container.
Pour the electrolyte into the electrolyte container. The level must
be between the minimum (350 ml) and maximum (750 ml)
marks.
Mount the container on the pump and mount the unit in the
cabinet of Movipol-3.
Before emptying the container of electrolyte, the section, Safety
Precautions should be carefully observed.
Remove the pump-container unit from the cabinet and remove
the container.
If the electrolyte is to be used again pour the electrolyte carefully
back into the electrolyte bottle. Use the funnel supplied with the
apparatus.
If the electrolyte is not to be used again it should be filled into a
container suitable for disposal. Please observe the local
regulations and instructions for disposal.
Funnel, gloves, ventilation as well as all other prescribed equipment must be
Press the polishing chamber onto the pencil until it fits tightly.
A polishing chamber has a limited lifetime as the plastic hardens
because of contact with the electrolyte. The chamber should be
renewed regularly.
IMPORTANT
used during filling/emptying.